Offshore Oil and Gas Calibration, Proving and Measurement Services and Solutions

Intertek KCS offers high volume meter proving (HVP) to the oil and gas industry along and in the Gulf of Mexico. Liquid measurement calibration simulation skids are maintained for accuracy and are in compliance with regulations set forth by the CFR’s and MPMS. The HVP simulates flow and proves accuracy of larger meters. The high volume dispacement prover is available for scheduled demonstration. KCS HVP meter proving experts bring solutions to your large meter proving requirements.

USA Homepage
Available Services
USA Services A-Z
Locations
Send Us A Request

High volume meter calibration

trace metal detection Prefer to call us?
Europe +44 1708.680.248 Americas:
  • Toll Free USA: 1 888 400 0084
  • All Others: +01 281 971 5600

  • Intertek KCS Homepage
  • Contact for information
  • Global E&P Services
  • Global Laboratory Capabilities
  • Intertek KCS Locations:

    Houma and Lake Arthur
    Louisiana USA

    Oil and Gas High Volume Meter Proving

    Kelley Completion Services

    Providing the Industry Standard for Proving High Volume Turbine Meters, with a maximum flow rate of 2750 GPM plus.

    Meter Proving services apply to:

    High Volume Meter Proving (HVP)

    High Volume Meter Calibration

    Turbine Meter Proving

    Turbine Meter Calibration

    Positive Displacement Meter Proving

    Positive Displacement Meter Calibration

    Liquid Meter Proving

    Liquid Meter Calibration

    The HVP helps to meet the calibration prover needs of the industry for increasing flow diameters while measuring flow accurately. Larger flow diameters challenge conventional meters to provide accurate and compliable liquid meter provings, while the HVP is capable of handling the higher flow volume requirements.

    High Volume Meter Prover System Capabilities:
    Medium: Diesel, Gravity: 33-34 API
    Medium Capacity: 650 bbl, Holding Tanks: 3 – 210 bbl
    Pump: 140 hp 460 v, Pump Size 10” X 8”
    System Bypass, Pulse Interpolation
    In-Line Flow Rate Detector, Proving Software
    Temperature Determination Transmitter
    Pressure Determination: Transmitter
    Meter Pulse Determination
    Four -Way Valve Actuation
    Meter Factor Determination
    Maximum Flow Rate: 2750 GPM Plus

    High Volume Prover Specifications:
    Prover Type: Bi-Directional, Serial #: 4102003
    Pipe ID: 15.25, ANSI: 300, Wall Thickness: .375
    Manufacturer: KCS, L.L.C.
    4-Way Valve: 8” General
    Mfg Switch: Magtek
    Coated and Covered
    Composition: Carbon Steele
    Round Trip Volume:
    Short Volume: 9.543036 bbl
    Long Volume: 29.208823
    Waterdraw Certified By: Meter Engineers, Inc.
    4-10-2003

    KCS High Volume Prover (HVP) System Capabilities:

    High Flow Rate: Rates through 2750 gpm (95,000 bbl/d) can be achieved through this system with prover sphere travel remaining at 5 feet per second. 95,000 bbl/d is 80% of the maximum flow rate of a standard 8” turbine meter. The high volume prover meets min and max ball speed allowances for bi-directional provers according to API-MPMS. Energy required to drive this system reaches 480 volts. KCS HVP meet these criteria and more.

    Superior Temperature Stabilization:
    Temperature fluctuation during proving runs are kept to a minimum using design considerations such as a by-pass system, long radius pipe bends, reduced friction, oversized pump, & increased retention time. The KCS HVP takes temperature stabilization into consideration in order to achieve accurate meter proving. The HVP reduces pump rotation against stagnant medium, reducing any increase in medium temperature.

    Large Volume Meter Proving and Calibration

    Medium Retention Time Increased:
    Longer medium retention time design allows the medium greater time to stabilize prior to being re-circulated back through the system. KCS has three 210 barrel retention storage tanks, allowing adequate storage to meet retention criteria.

    Minimum Pulses Capability:
    According to API-MPMS Chapter 4 sec. 2, minimum required pulses between detector switches are 10,000. On standard turbine meters the HVP can obtain these required pulses through 4” meters. Minimum required pulses between detector switches are 10,000, per API-MPMS Chapter 4 Section 2 .

    Double Chronometry Pulse Interpolation:
    For those meters whose discrimination does not allow the HVP to obtain 10,000 pulses between detectors, a Double Chronometry software package is utilized. This enables the HVP to increase the readout discrimination of the flow-meter pulses to achieve an uncertainty of 0.01% according to API-MPMS Chapter 4 sec 6 Pulse Interpolation. KCS HVP gives adequate flow rates, stabilized temperature, and sufficient pulses in compliance when proving larger meters on the HVP simulation skid.